DM32 Gauge

See how you can check your gauge’s calibration against itself and another gauge. Learn a simple way to test positive and negative ports against each other.https://youtu.be/nB7GbaG-aDI

How It Works: Step-by-Step

Set Both Channels to Pressure Mode

Switch both Channel A and Channel B to measure pressure in Pascals (Pa).

Cross-Port Input Comparison

Connect a tube between the Input ports of both channels (blue and green).
Both channels should display the same pressure within 2% accuracy.
(Blue = Input A; Green = Input B)

Cross-Port Reference Comparison

Disconnect and reconnect the tube between the Reference ports (red and yellow).
Again, readings must match within 2%.
(Red = Ref A; Yellow = Ref B)

Compare Against Another Calibrated Gauge (Optional)

For added confidence, connect corresponding ports between your gauge and a known-nearby calibrated gauge.
Readings should also align within 2%.

Backed by Manufacturer Guidelines

Retrotec recommends this quick check as part of weekly or pre-test routines. It identifies gauge drift, blocked ports, or tubing issues rapidly—without needing to send the gauge for lab calibration—since most field errors aren’t calibration-related but due to tube faults or port mismatches.

Quick Checklist

ModePortsActionExpected Outcome
InputBlue–GreenConnect tubeChannels match within 2%
RefRed–YellowConnect tubeChannels match within 2%
BothGauge–GaugeCompare same ports across two gaugesWithin 2%

If any comparison fails:

  • Inspect and replace tubing

  • Test alternate ports

  • Only consider factory calibration if all tubing and ports pass but results still differ